Skip to content

Passeo-14
Peripheral Dilatation Catheter

Indicated for balloon dilation of the stenotic portion of a lower limb artery for the purpose of improving perfusion.*

Technical Data

Balloon Catheter
Catheter typeOTW
Recommended guide wire0.014”
TipOptimized entry profile and colored
Balloon materialSCP (Semi-Crystalline Polymer), controlled compliance (4 - 6%)
Balloon folding3-fold
Balloon coatingHydrophilic patchwork coating
Balloon markers2 swaged markers (zero profile)
Sizes1.5 - 4.0 mm; L: 20 - 220 mm
Distal shaft3.1F, hydrophilic coating, coaxial design; 150 mm length (1.5 /2.0 x 20 - 100 mm); 75 mm length (2.0 x 140 - 220 mm and 2.5 - 4.0 mm)
Proximal shaft3.9F, hydrophobic coating, coaxial design; stiffening wire
Usable length150 cm (1.5 - 4.0 mm); 120 cm (1.5 - 2.0 mm); 90 cm (2.5 - 4.0 mm)

Compliance Chart

Balloon diameter x length (mm)
1.5 x 20-702.0 x 40-2202.5 x 40-2203.0 x 40-2203.5 x 40-1404.0 x 40-140
Nominal Pressure (NP)atm**777777
(mm)1.52.02.53.03.54.0
Rated burst Pressure (RBP)atm**141414141414
(mm)1.572.082.613.183.634.16
**1 atm = 1.013 bar

Ordering Information

Catheter Length (cm)Balloon (mm)Balloon Length (mm)
204070100140180220
Antegrade approach4F1201.5380271380277380283----
Antegrade approach4F1202.0-380278380284380290380296380302380308
Antegrade approach4F902.5-380279380285380291380297380303380309
Antegrade approach4F903.0-380280380286380292380298380304380310
Antegrade approach4F903.5-380281380287380293380299-
Antegrade approach4F904.0-380282380288380294380300-
Catheter Length (cm)Balloon (mm)Balloon Length (mm)
204070100140180220
Crossover approach4F1501.5380313380319380325----
Crossover approach4F1502.0-380320380326380332380338380344380350
Crossover approach4F1502.5-380321380327380333380339380345380351
Crossover approach4F1503.0-380322380328380334380340380346380352
Crossover approach4F1503.5-380323380329380335380341--
Crossover approach4F1504.0-380324380330380336380342--