
©2021 Vascular Solutions LLC 1 103805 Rev I 01/23
English/Instructions for Use ........................................1
Vascular Solutions LLC
6464 Sycamore Court North
Minneapolis, MN 55369 USA
(866) 246-6990
(866) 804-9881
Wattson™ temporary pacing guidewire
Instructions For Use
USA CAUTION
Federal law restricts this device to sale by or on the order of a
physician.
DEVICE DESCRIPTION
The Wattson temporary pacing guidewire is a dual-purpose 0.035”
guidewire designed for the delivery of devices and for temporary
pacing of the heart. The distal end of the device has an atraumatic
pigtail shape. The shaft of the device has a lubricious hydrophilic
coating.
Pacing is achieved through a bipolar electrode conguration
integrated in the shaft of the guidewire. The electrodes terminate
in a single positive electrode at the distal tip of the device and
three discrete negative electrodes located proximal to the tip within
the pigtail section. The entire device is visible using standard
uoroscopic methods.
The device is packaged with a proprietary adapter that locks on to
the proximal end of the guidewire and converts it to discrete positive
and negative terminal pins. The terminal pins of the adapter are
compatible with standard external pulse generators.
The Wattson temporary pacing guidewire has been sterilized with
ethylene oxide.
SPECIFICATIONS
Model 2250
Length 280cm
Shaft outer
diameter
0.035” (0.89mm)
Distal
pigtail outer
diameter
30mm
Lubricious
Hydrophilic
Coating
80cm
280 cm
0.035”
(0.89 mm)
30 mm
80 cm
Figure 1: Wattson™ temporary pacing guidewire dimensions
Figure 2: Wattson™ temporary pacing guidewire with adapter
assembly and adaptor pins
INDICATIONS
The Wattson temporary pacing guidewire is intended to introduce
and position catheters and other interventional devices within the
chambers of the heart, including those used within transcatheter
aortic valve replacement (TAVR) procedures and balloon aortic
valvuloplasty (BAV), while transmitting an electrical signal from
an external pulse generator to the heart. The temporary pacing
guidewire is not intended to remain in place following the clinical
procedure.
CONTRAINDICATIONS
The guidewire is contraindicated for use in the coronary arteries and
in the cerebrovasculature.
WARNINGS
Failure to abide by the warnings in this labeling might result in
damage to the device coating, which may necessitate intervention or
result in a serious adverse event.
For patients at high risk of developing new highgrade AV conduction
block during or immediately following the index procedure, the
Wattson temporary pacing guidewire should not constitute the sole
source of ventricular pacing.
The Wattson temporary pacing guidewire is provided sterile for
single use only. Do not reuse or resterilize. Reuse of a single-use
device creates a potential risk of patient or user infections and may
compromise device functionality, which may lead to illness or serious
patient injury.
Never advance or withdraw an intravascular device against
resistance until the cause of the resistance is determined by
uoroscopy. Movement of the guidewire against resistance may
result in separation of the guidewire tip, other damage to the
guidewire, coating material remaining in the vasculature or patient
injury requiring additional intervention.
Do not torque the device, as tip separation or patient injury may
occur.
Inappropriate electrical connections (e.g., into a wall socket) may
pose serious risk of adverse health consequences.
Do not use the guidewire with magnetic resonance imaging (MRI).
PRECAUTIONS
The safety and eectiveness of the coated device has not been
established, or is unknown, in vascular regions other than those
specically indicated.
Do not use the Wattson temporary pacing guidewire if the packaging
is damaged. A damaged package could indicate a breach of sterility
or device damage.
Exercise care in handling the guidewire during a procedure to reduce
the possibility of accidental breakage or kinking.
When the guidewire is in the body, it should be manipulated only
under uoroscopy. Do not attempt to move the guidewire without
observing the resultant tip response, as guidewire damage or vessel
injury may occur.
Do not reshape the tip of the guidewire, as tip separation may occur.
Always secure the guidewire prior to advancing or withdrawing an
interventional device over it. Failure to do so may result in loss of
guidewire position and/or vessel injury.
Do not withdraw or manipulate the guidewire in a metal cannula,
needle, or other sharp-edged instrument, as it may impair the
integrity of the polymer coating and compromise the trackability of
the guidewire.
Do not insert other objects into the adapter unless it is the proximal
end of the Wattson temporary pacing guidewire. Inserting other
guidewires or other devices into the adapter may result in adapter
damage.
Care must be taken not to remove the Wattson temporary pacing
guidewire from the adapter or move the guidewire during pacing as
loss of capture or inadequate pacing may result.
ADVERSE EFFECTS
Potential adverse eects that may be associated with the temporary
pacing guidewire include, but are not limited to, the following:
• Induction of non-intended arrhythmia
• Dissection
• Embolism
• Hematoma
• Hypotension
• Infection
• Left ventricular perforation
• Stroke
• Tamponade
• Thrombosis
• Sterile inammation or granulomas at the access site
• Additional surgical intervention
CLINICAL PROCEDURE
The Wattson temporary pacing guidewire should be used by
physicians trained on the procedures for which the device is
intended. The techniques and procedures described do not
represent ALL medically acceptable protocols, nor are they intended
as a substitute for the physician’s experience and judgment in
treating any specic patient. All available data, including the patient’s
signs and symptoms and other diagnostic test results, should be
considered before determining a specic treatment plan.
Package contains:
• 1x Guidewire
• 1x Adapter
• 2x Adapter pins
Other items required but not provided:
• External pulse generator
PREPARATIONS FOR USE
1. Prior to use, carefully inspect the guidewire packaging and
components for damage.
2. Utilizing sterile technique, remove the guidewire hoop and
adapter from the packaging and transfer them both to the
sterile eld.